Cosmic Ray (Muon) Detector using Geiger–Müller Tubes

This detector is to be used as test unit, to measure the performance of my other project using Fluorescent Tubes against in order to clarify and identify any issues, also to better understand and to also demonstrate the principles of a cosmic ray telescope.

Audio Recording of the Detector Outputs (1.35 MB Mp3)

Above is one of the first visual test of the circuit, which has been updated a number of times since this video. Although the new circuit produces less false positives and hence less flashes, the video here offers a good demonstration of what the output data looks like. The LED flash times are slowed by a one-shot timer, as the pulses are so short they would not be visible if the LEDs where driven directly.

 


Version 7. 1st August 2009

 

After a number of different configurations and tests, I have distilled the design down to a simple circuit seen here. The outputs 4,5,6 give a positive 5V logic pulse when a coincidence is detected in two or more of the tubes.

 

 

Although Geiger–Müller tubes are sensitive to Muons, the response time to decay (Pulse Width) when a muon is detected is relatively long for measuring the probability of coincidence in two or more the tubes. This means the wider the Pulse Width the greater the number of false positives. Consequently some means of pulse shortening is required to shorten the Pulse Width to a shorter period to decrease the probability that it is terrestrial radiation hitting the tubes in close succession.

 

 

The CRO trace above demonstrates what happens in this circuit when a wider pulse in feed in. The bottom trace is the output and only responds to a negative travelling pulse regardless of pulse shape. This is also important as the Geiger–Müller tube is positively biased and when a particle is detected the output swings negative, so the circuit ensures that only the first micro second of the detector pulse is processed in the coincidence circuit.

 

 

Why 3 tubes? I hear you say, well it is all about increasing the detectors aperture size to muons as most Geiger–Müller tubes available are longer than they are wide. So as an attempt to achieve a larger aperture as well as coincidence detection without using rows of tubes in each layer I am combining the coincidence outputs from top-middle and middle-bottom detectors.

 

 

The prototype provides six positive 5V logic outputs to a din socket. 1) Top tube - all detections 2) Middle tube - all detections 3) Bottom tube - all detections (all detections) meaning both cosmic and background radiation 4) Top and Middle - coincidence detection 5) Bottom and Middle - coincidence detection 6) "Top & Middle" or "Middle & Bottom" - coincidence detection (coincidence detection) meaning a stronger likelihood of a cosmic source than terrestrial source being detected. Outputs 4 and 5 are the main outputs, but to give some visual information that the detector is working I've added some LEDs driven by one-shot timers to give a 1/8 second flash, as the output pulses are only a micro second and wouldn't be seen if used to drive the LEDs directly. Output 6 drives the blue LED indicating a confirmed coincidence in two or more tubes and where outputs 1,2 & 3 drive the red LEDs indicating any detection in the adjacent tube.

 

 

Coincidence Detection The main idea of 'coincidence detection' in signal processing is that if a detector detects a signal pulse in the midst of random noise pulses inherent in the detector, there is a certain probability, p, that the detected pulse is actually a noise pulse. But if two detectors detect the signal pulse simultaneously, the probability that it is a noise pulse in the detectors is p2. Suppose p = 0.1. Then p2 = 0.01. Thus the chance of a false detection is reduced by the use of coincidence detection.

 

Detector Using 2 Pancake Geiger–Müller Tubes

I'm also making a two tube variation of the 3 tube design using Pancake Geiger–Müller tubes as they have a larger surface area.

Issues with detectors using Geiger–Müller Tubes

What should the count rate be?
The theoretical rate of cosmic rays is of the order of 1 count per minute per cm2 of active area, but it does depend on solid angle, so this number is only approximate. Assuming each GM tube has a broadside active area about 10 cm2, the number of counts maybe 10 per minute.

If the rate of each tube; S1, S2, and S3 counts per second and the coincidence gate width is τ seconds, then if counter #1 is ON S1τ , and the random coincidences in tube #2 at a rate R12 = S1S2τ random coincidences per second. So the random coincidence rate should be less than 1% of what is expected for the real coincidence rate, or about 0.1 counts per minute for a 10 cm2 Detector. (Many thanks to Bob S for this information)

Compton Scattering
One of the reasons for false counts in a Geiger–Müller array detector maybe due to Compton Scattering, where an interaction between charged electrons within the detector and high energy photons result in the electron being given part of the energy, causing a recoil effect of a high energy photon into the adjacent detector causing a false coincidence detection. In other words this causes cross-talk interference between GM Tubes

Consequently radiation shielding is required between each GM Tube of either 6mm of lead, 12mm of copper or 25mm of aluminium (note Iron is unsuitable). (Many thanks to Bob S for this information)

August 13th 2009 - To date tests carried out using radiation shielding between the GM tubes don't indicate that this is a real problem, however any cross-talk between each the tubes or the electronic is a real concern and should be design into any detector array.

Detector pulse width
In theory the Geiger–Müller tube is a very good detector of Muons (Cosmic Rays) however it would seem that filtering out background radiation using a simple coincidence detector is problematic due to the Geiger–Müller tube response and decay time (Pulse Width) when an ionising particle has been detected.

Consequently, the wider the Pulse Width the greater the number of false positives. The means a pulse shorting or quenching circuit is also needed to shorten the Pulse Width to a period closer to the expected flight time of the Muon between tubes, but not too narrow that the electronics cannot measure relative coincidence. Some improvement might also be achieved by spacing the tubes further apart, but this also has the negative effect of decreasing the aperture of the detector.

Coincidence
Although coincidence implies simultaneously, in reality we are talking about almost simultaneous, this is because a muon created by a cosmic event is travelling at near the speed of light 0.998c, so if the detectors are only spaced 2.5cm apart the actual flight time of a muon would only be 0.08ns. However as the detector and electronics response and delay times are slower than this, we can say in "real-life terms" it is simultaneous.

Sonification of Muons via Arduino and Max/MSP

I loaned my 3 Tube Muon Detector to Sebastian Tomczak who has rigged it up for some experimental sonification.